60601-1 edition 3.1 and understanding IEC60601-1 Document structure

Download PDF

Globtek applies for IEC 60601-1 medical Safety Approvals and EMC approvals on their standard power supply products. The medical approvals for Safety and EMC are required to exist in separate CB reports.

Many medical products are used in a home health care environment. Globtek obtains a IEC 60601-1-11 approval on their products in order to support this important market segment of the medical equipment business. An important distinction with home use medical equipment is that the AC input configuration must be a Class II (unearthed, 2 conductor) configuration.

There is often confusion regarding the differences between IEC 60601-1 ed 3, and IEC 60601-1 ed 3.1. As shown in the diagram below, when Corrigendum 1 and 2, and Amendment 1 are applied to edition 3, the resulting document is the same as edition 3.1. The changes are generally minor between ed 3 and ed 3.1, one major change is to have relaxed safety isolation barrier requirements for low voltage DC systems, such as automobile batteriesl, where the risk is far less than a 230Vac input safety barrier.

In the diagram below, the two key collateral standards for IEC 60601-1 which Globtek obtains CB approvals (and specific harmonized national standards approvals) for are shown, which is for the medical EMC standard and the Home Use Care standard.

diagram

For additional information on this topic, please visit the following web pages: