GT-41060-WWVV-TAB Wall Plug-in ICT / ITE / Medical Power Supply

GT-41060-WWVV-TAB
GT-41060-WWVV-TAB, ICT / ITE / Medical Power Supply, Wall Plug-in, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, NEMA 1-15P, North America Blades, Class II 2 Conductors, Output Rating: 25 Watts, Power rating with convection cooling (W) , 3.0-30V in 0.1V increments, Approvals: Belarus; SIQ; CE; China RoHS; Double Insulation; FCC; Level IV 1meter 115; RoHS; Ukraine; VCCI; WEEE; CB 60950; Book 60601; PSE; 2MOPP; cULus;GT-41060-WWVV-TAB, ICT / ITE / Medical Power Supply, Wall Plug-in, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, NEMA 1-15P, North America Blades, Class II 2 Conductors, Output Rating: 25 Watts, Power rating with convection cooling (W) , 3.0-30V in 0.1V increments, Approvals: Belarus; SIQ; CE; China RoHS; Double Insulation; FCC; Level IV 1meter 115; RoHS; Ukraine; VCCI; WEEE; CB 60950; Book 60601; PSE; 2MOPP; cULus;
Information
Model NumberGT-41060-WWVV-TAB
Agency Documentshttp://www.globtek.info/certs/GT-41060/
CE EC-Declaration for MedicalContact GlobTek for more information
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0Oa000000OJ1ESEA1
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for MedicalContact GlobTek for more information
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000OJ1ESEA1
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000OJ1ESEA1
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0Oa000000OJ1ESEA1
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeWall Plug-in
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)0.6 A
Max Output Power Rating (W)25
Output Voltage Range (V)3.0-30
EfficiencyIV
Ingress ProtectionIP40
Size (mm)
EnclosureUser-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)Output NumberVoltageAmps (Convection)Amps (Forced Air)Watts (Convection)Watts (Forced Air)RFQ
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS:
1. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac
2. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz
3. Output Regulation: +/- 5% measured at the output connector
4. Line Voltage Regulation: +/- 1% maximum at full load
5. Output Ripple (Vp-p): 1% or 150 mV whichever is greater measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of output connector at nominal line
6. Turn-On/Turn-Off Overshoot: 5% maximum, 500uS maximun recovery time for 25% to 50% step load
7. Turn-On Delay: 3000 mSEC maximum at full load, 115VAC
8. Hold-Up Time: 10mS minimum at nominal input and full load
9. Inrush Current: 30A typical at 115VAC input and 60A typical at 230 VAC input
10. Switching Frequency: 65 KHz typical

B) PROTECTION
1. Over-Voltage: Protected, unit will recover upon removal of fault
2. Over-Load: Protected, unit will recover upon removal of fault
2. Short Circuit: Protected, unit will recover upon removal of fault
3. Input Protection: Input line fusing

C) SAFETY
1. Dielectric Withstand Voltage: 4242Vdc from primary to secondary
2. Touch Current: <0.25mA @ 240Vac input voltage

D) OTHER:
1. MTBF: 200,000 hours @ 25°C ambient temperature
2. Operating Temperature: 0°C to 40°C ambient temperature
3. Humidity: 0% to 90% relative humidity
4. Storage Temperature: -40°C to 80°C
5. ROHS 2: Complies with EU 2011/65/EU & CHINA SJ/T 11363-2006

E) ENCLOSURE
1. Upper Housing: High impact plastic, 94V0 polycarbonate, non-vented
2. Lower Housing: High impact plastic, 94V0 polycarbonate, non-vented
3. Size: 84.0* 56.0*37.9 mm +/-1.0
4. Green LED indicator for power "Power On"
Derating Curve
Qty (ea.)/Carton40
Individual Box/Gift Box Size (mm)114*69.5*58.4
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)463 x 355 x 122
Gross Weight (kg)10.8
Label Width (mm)48.5
Label Height (mm)40.8
Input Configuration
DescriptionNEMA 1-15P, North America Blades, Class II 2 ConductorsAll NEMA 1 devices are two-wire non-grounding devices (hot-neutral) rated for 125 V maximum. NEMA 1-15P plugs have two parallel flat blades, 0.25 inches (6.35 mm) wide, 0.06 inches (1.524 mm) thick, 0.625-0.718 inches (15.875–18.256 mm) long, and spaced 0.5 inches (12.7 mm) apart.
Output Configuration
Common output connector options:
Click to add to your request.
L Type (Coaxial 5.5x2.5mm plug)
C Type (Coaxial 5.5x2.1mm plug)
K Type (Coaxial 3.5x1.3mm plug)
LL Type (5.5x2.5mm Locking 760k type)
CL Type (5.5x2.1mm Locking S761k type)
Magnetic Connector
L Type (Coaxial 5.5x2.5mm plug)C Type (Coaxial 5.5x2.1mm plug)K Type (Coaxial 3.5x1.3mm plug)LL Type (5.5x2.5mm Locking 760k type)CL Type (5.5x2.1mm Locking S761k type)Magnetic Connector
YL3 Type (KPPX-3P)
YL4 Type (KPPX-4P)
EJ1/2/3/4/5 (EIAJ RC-5320A type connectors)
MSB Type (Micro USB)
Custom design
YL3 Type (KPPX-3P)YL4 Type (KPPX-4P)EJ1/2/3/4/5 (EIAJ RC-5320A type connectors)MSB Type (Micro USB)Inquire for custom design

Comprehensive list of GlobTek output cordset options

Contact GlobTek for your specific requirements or custom solutions.

Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.