GTM962253PWWWVV-P3 Potted ICT / ITE / Medical Power Supply

GTM962253PWWWVV-P3
GTM962253PWWWVV-P3, ICT / ITE / Medical Power Supply,  60601-1-4th Ed. , Potted, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Contact GlobTek for customization requests on any type of cable in any customer specified lengths., Output Rating: 225 Watts, Power rating with convection cooling (W) , 12-54V in 0.1V increments, Approvals: IP68;GTM962253PWWWVV-P3, ICT / ITE / Medical Power Supply, 60601-1-4th Ed. , Potted, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Contact GlobTek for customization requests on any type of cable in any customer specified lengths., Output Rating: 225 Watts, Power rating with convection cooling (W) , 12-54V in 0.1V increments, Approvals: IP68;
Information
Model NumberGTM962253PWWWVV-P3
Agency Documentshttp://www.globtek.info/certs/GTM96225xP%20/
CE EC-Declaration for Medicalhttps://cs.globtek.info/ec_declaration?id=a0O3a00000MlVt2EAF
CE EC-Declaration for Industrial/Commercial/ITEContact GlobTek for more information
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://cs.globtek.info/ukca_declaration?id=a0O3a00000MlVt2EAF
UKCA-Declaration for Industrial/Commercial/ITEContact GlobTek for more information
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0O3a00000MlVt2EAF
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O3a00000MlVt2EAF
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypePotted
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)3A
Max Output Power Rating (W)225
Output Voltage Range (V)12-54
EfficiencyUSA DOE Level VI / Eco-design Directive 2009/125/EC, (EU) 2019/1782
Ingress ProtectionIP68
Size (mm)181.0 (L) x 70.0(W) x 39.0 (H)
Enclosure
User-added image

The black outlined drawing represents a housing which is being phased out due to end of usable life of the mold used to make it. The blue outlined drawing is the replacement housing. Both may be used interchangeably and mixed within any given lot. 
3D File Downloadhttps://www.globtek.com/file/dl/GTM962253PWWWVV-P3.zip
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTM962253P14016-P3*168.75140
GTM962253P14018-P3*187.77139.86
GTM962253P14019-P3*197.36139.84
GTM962253P14022-P3*226.36139.92
GTM962253P14024-P3*245.83139.92
GTM962253P14030-P3*304.66139.8
GTM962253P14048-P3*482.91139.68
GTM962253P14015-P3*159.33139.95
GTM962253P14036-P3*363.88139.68
GTM962253P14028-P3*285140
GTM962253P14054-P3*542.59139.86
GTM962253P08412-P3*12784
GTM962253P14020-P3*207140
GTM962253P14012-P3*1211.66139.92
GTM962253P14016-P3*168.75140
GTM962253P14018-P3*187.77139.86
GTM962253P14019-P3*197.36139.84
GTM962253P14022-P3*226.36139.92
GTM962253P14024-P3*245.83139.92
GTM962253P14030-P3*304.66139.8
GTM962253P14048-P3*482.91139.68
GTM962253P14015-P3*159.33139.95
GTM962253P14036-P3*363.88139.68
GTM962253P14028-P3*285140
GTM962253P14054-P3*542.59139.86
GTM962253P08412-P3*12784
GTM962253P14020-P3*207140
GTM962253P14012-P3*1211.66139.92
GTM962253P14016-P3*168.75140
GTM962253P14018-P3*187.77139.86
GTM962253P14019-P3*197.36139.84
GTM962253P14022-P3*226.36139.92
GTM962253P14024-P3*245.83139.92
GTM962253P14030-P3*304.66139.8
GTM962253P14048-P3*482.91139.68
GTM962253P14015-P3*159.33139.95
GTM962253P14036-P3*363.88139.68
GTM962253P14028-P3*285140
GTM962253P14054-P3*542.59139.86
GTM962253P08412-P3*12784
GTM962253P14020-P3*207140
GTM962253P14012-P3*1211.66139.92
GTM962253P14016-P3*168.75140
GTM962253P14018-P3*187.77139.86
GTM962253P14019-P3*197.36139.84
GTM962253P14022-P3*226.36139.92
GTM962253P14024-P3*245.83139.92
GTM962253P14030-P3*304.66139.8
GTM962253P14048-P3*482.91139.68
GTM962253P14015-P3*159.33139.95
GTM962253P14036-P3*363.88139.68
GTM962253P14028-P3*285140
GTM962253P14054-P3*542.59139.86
GTM962253P08412-P3*12784
GTM962253P14020-P3*207140
GTM962253P14012-P3*1211.66139.92
GTM962253P14016-P3*168.75140
GTM962253P14018-P3*187.77139.86
GTM962253P14019-P3*197.36139.84
GTM962253P14022-P3*226.36139.92
GTM962253P14024-P3*245.83139.92
GTM962253P14030-P3*304.66139.8
GTM962253P14048-P3*482.91139.68
GTM962253P14015-P3*159.33139.95
GTM962253P14036-P3*363.88139.68
GTM962253P14028-P3*285140
GTM962253P14054-P3*542.59139.86
GTM962253P08412-P3*12784
GTM962253P14020-P3*207140
GTM962253P14012-P3*1211.66139.92
GTM962253P14016-P3*168.75140
GTM962253P14018-P3*187.77139.86
GTM962253P14019-P3*197.36139.84
GTM962253P14022-P3*226.36139.92
GTM962253P14024-P3*245.83139.92
GTM962253P14030-P3*304.66139.8
GTM962253P14048-P3*482.91139.68
GTM962253P14015-P3*159.33139.95
GTM962253P14036-P3*363.88139.68
GTM962253P14028-P3*285140
GTM962253P14054-P3*542.59139.86
GTM962253P08412-P3*12784
GTM962253P14020-P3*207140
GTM962253P14012-P3*1211.66139.92
GTM962253P14016-P3*168.75140
GTM962253P14018-P3*187.77139.86
GTM962253P14019-P3*197.36139.84
GTM962253P14022-P3*226.36139.92
GTM962253P14024-P3*245.83139.92
GTM962253P14030-P3*304.66139.8
GTM962253P14048-P3*482.91139.68
GTM962253P14015-P3*159.33139.95
GTM962253P14036-P3*363.88139.68
GTM962253P14028-P3*285140
GTM962253P14054-P3*542.59139.86
GTM962253P08412-P3*12784
GTM962253P14020-P3*207140
GTM962253P14012-P3*1211.66139.92
GTM962253P14016-P3*168.75140
GTM962253P14018-P3*187.77139.86
GTM962253P14019-P3*197.36139.84
GTM962253P14022-P3*226.36139.92
GTM962253P14024-P3*245.83139.92
GTM962253P14030-P3*304.66139.8
GTM962253P14048-P3*482.91139.68
GTM962253P14015-P3*159.33139.95
GTM962253P14036-P3*363.88139.68
GTM962253P14028-P3*285140
GTM962253P14054-P3*542.59139.86
GTM962253P08412-P3*12784
GTM962253P14020-P3*207140
GTM962253P14012-P3*1211.66139.92
GTM962253P14016-P3*168.75140
GTM962253P14018-P3*187.77139.86
GTM962253P14019-P3*197.36139.84
GTM962253P14022-P3*226.36139.92
GTM962253P14024-P3*245.83139.92
GTM962253P14030-P3*304.66139.8
GTM962253P14048-P3*482.91139.68
GTM962253P14015-P3*159.33139.95
GTM962253P14036-P3*363.88139.68
GTM962253P14028-P3*285140
GTM962253P14054-P3*542.59139.86
GTM962253P08412-P3*12784
GTM962253P14020-P3*207140
GTM962253P14012-P3*1211.66139.92
* (Final ordering #/Part No to be defined)
Specifications

Features

  • High power density potted construction
  • Low conducted & radiated EMI: Class B without external filters or ferrites (all input configurations, into resistive load)
  • Class I, Class II, or Class II with Functional Earth input configurations
  • ITE/ICT and medical (2 x MOPP) certified (suitable for Type-B or Type-BF applications)
  • Auxiliary +12V output

Input

Input Voltage:Specified: 85-264VAC, Nameplate: 100-240VAC
Input Frequency:Specified: 47-63Hz, Nameplate: 50-60Hz
Input Current:2.75A typ. @ 85VAC, 1.0A typ. @ 230VAC (full load)
No Load Input Power:<150mW
Inrush Current:<75A @ 230VAC (cold start @ 25°C)
Efficiency:90 - 95% typ. (full load)
Power Factor:>0.90 @ 115VAC, >0.85 @ 230VAC, 140W output

Output

Output Power:140W continuous
Turn-on Delay:< 1 second @ 85VAC
Output Voltage Tolerance:± 2.0% (worst case, set point + load/line regulation)
Load Regulation:± 0.5% (measured at connector)
Line Regulation:± 0.2% (measured at connector)
Ripple:1% pk-pk (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW)
Transient Response:<5% deviation (with 25% load step), recovery to within 1% in 1ms
Startup Overshoot:<5% (full load)
Hold-up Time:10ms min. (full load)
Aux Output Voltage Regulation:12V ± 2V (10V min. / 14V max.), for 0A ≤ IAux 0.8A

Aux regulation guaranteed when main output power ≥ 25% of 140W rating (~34W).

For main output power < 25%, aux output voltage may drop to < 50% of nominal rating, depending on aux output load. Please check with your system.
For main output = 12V, aux output voltage is same as main output.
Aux Output Current:0.8A

Protections

Input Protection:MOV transient suppressor, dual fuses (line + neutral)
Over-Voltage Protection:110-130%, latched-off, cycle AC to reset
Over-Current Protection:130% typical, auto-recovery
Short Circuit Protection:Auto-recovery
Over-Temperature Protection:NTC temperature sensor embedded in transformer, trip between 115 - 120°, latched-off, cycle AC to reset (auto-recovery available as special option)

Environmental

MTBF:1,500,000 hours @ 25°C ambient, full load (Telcordia SR-332, Issue 3)
Operating Temperature:-20°C to 50°C (full output power), -25°C to 80°C (de-rated)
Storage Temperature:-30°C to 85°C
Humidity:0% to 95% relative humidity, non-condensing
Altitude:5000m max.
Vibration:Conforms to EN60068-2-6, ISO80601-2, EN1789

Safety

Dielectric Withstand Voltage:4000VAC or 5656VDC from input to output
3000VAC or 4242VDC from input to earth
Intended Medical Application:Type-B or Type-BF
Touch Current:3-conductor input - NC: <25µA, SFC: <200µA
2-conductor input - NC: <65µA, SFC: <140µA
Earth Leakage Current3-conductor input: <200µA
2-conductor input: N/A
Means of Protection:2 x MOPP
Primary-Secondary Bridging Capacitance:Two IEC60384-14 certified Y1 capacitors in series

EMC

Applicable Standards:Medical: EN60601-1-2, 4th edition
Emissions: EN55032, EN 55011, EN61000-6-3, EN61000-6-4, CISPR11/CISPR22
Immunity: EN55024, EN61000-6-1, EN61000-6-2
Conducted Emissions:Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:EN61000-3-2, Class A
Voltage Fluctuations/Flicker:EN61000-3-3
Electrostatic Discharge (ESD) Immunity:EN61000-4-2, 10KV contact discharge, 20KV air discharge (applied to input or output terminals)
Radiated RF Immunity:EN61000-4-3, 20V/m 80-2700MHz, 80% 1KHz AM, 80% 5Hz AM
EFT/Burst Immunity:EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:EN61000-4-6, 3VRMS, (6VRMS in ISM/amateur bands), 80% 1KHz AM
Power Frequency Magnetic Field Immunity:EN61000-4-8, 30A/m
Voltage Dip Immunity:EN61000-4-11
 

Special Options

  • Custom labels/markings
  • Custom input/output cable assemblies
  • Short term output surge capability
  • Reduced touch current
  • Reduced output ripple
  • Compliance to IPC-600, IPC-610, IPC-6012 Class 2 or 3
  • IEC61000-3-2 Class-C line harmonics for LED lighting applications
  • Non-latching (auto-recover) over-temperature protection
  • Potentiometer to adjust/trim output voltage
  • Custom aux output voltages (5V to 12V), available for main output voltages from 12-24V

 

Derating CurveUser-added image
Qty (ea.)/Carton20
Individual Box/Gift Box Size (mm)Cellular Packag
Individual Box/Gift Box Color
Master Carton Size (mm)450 x 461 x 182
Gross Weight (kg)16
Label Width (mm)TBD
Label Height (mm)TBD
Input Configuration
DescriptionContact GlobTek for customization requests on any type of cable in any customer specified lengths.
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Medical 4th Ed.EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
IP68Ingress Protection: ?IP68 to IEC60529:2001 Totally dust tight and Protected against long periods of immersion under pressure
IP68Ingress Protection: ?IP68 to IEC60529:2001 Totally dust tight and Protected against long periods of immersion under pressure
IP68Ingress Protection: ?IP68 to IEC60529:2001 Totally dust tight and Protected against long periods of immersion under pressure
IP68Ingress Protection: ?IP68 to IEC60529:2001 Totally dust tight and Protected against long periods of immersion under pressure
IP68Ingress Protection: ?IP68 to IEC60529:2001 Totally dust tight and Protected against long periods of immersion under pressure
IP68Ingress Protection: ?IP68 to IEC60529:2001 Totally dust tight and Protected against long periods of immersion under pressure
IP68Ingress Protection: ?IP68 to IEC60529:2001 Totally dust tight and Protected against long periods of immersion under pressure
IP68Ingress Protection: ?IP68 to IEC60529:2001 Totally dust tight and Protected against long periods of immersion under pressure
IP68Ingress Protection: ?IP68 to IEC60529:2001 Totally dust tight and Protected against long periods of immersion under pressure

Request Info