Preferred ratings shown below. Custom output voltages available in 0.1V increment from the minimum voltage listed to the maximum voltage listed; contact sales.
Model Number (Final ordering #/Part No to be defined)
Voltage (V)
Current (A)
Power (W)
RFQ
GTM961200P11012-T3A*
12
9.2
110.4
GTM961200P11013-T3A*
13
8.54
111.02
GTM961200P11014-T3A*
14
7.93
111.02
GTM961200P12015-T3A*
15
8
120
GTM961200P12016-T3A*
16
7.5
120
GTM961200P12017-T3A*
17
7.06
120.02
GTM961200P12018-T3A*
18
6.66
119.88
GTM961200P12019-T3A*
19
6.316
120
GTM961200P12020-T3A*
20
6
120
GTM961200P12021-T3A*
21
5.71
119.91
GTM961200P12022-T3A*
22
5.45
119.9
GTM961200P12023-T3A*
23
5.21
119.83
GTM961200P12024-T3A*
24
5
120
GTM961200P12025-T3A*
25
4.8
120
GTM961200P12026-T3A*
26
4.61
119.86
GTM961200P12027-T3A*
27
4.44
119.88
GTM961200P12028-T3A*
28
4.28
119.84
GTM961200P12029-T3A*
29
4.13
119.77
GTM961200P12030-T3A*
30
4
120
GTM961200P12031-T3A*
31
3.87
119.97
GTM961200P12032-T3A*
32
3.75
120
GTM961200P12033-T3A*
33
3.63
119.79
GTM961200P12034-T3A*
34
3.53
120.02
GTM961200P12035-T3A*
35
3.42
119.7
GTM961200P12036-T3A*
36
3.33
119.88
GTM961200P12037-T3A*
37
3.24
119.88
GTM961200P12038-T3A*
38
3.15
119.7
GTM961200P12039-T3A*
39
3.07
119.73
GTM961200P12040-T3A*
40
3
120
GTM961200P12041-T3A*
41
2.92
119.72
GTM961200P12042-T3A*
42
2.85
119.7
GTM961200P12043-T3A*
43
2.79
119.97
GTM961200P12044-T3A*
44
2.72
119.68
GTM961200P12045-T3A*
45
2.66
119.7
GTM961200P12046-T3A*
46
2.6
119.6
GTM961200P12047-T3A*
47
2.55
119.85
GTM961200P12048-T3A*
48
2.5
120
GTM961200P12049-T3A*
49
2.44
119.56
GTM961200P12050-T3A*
50
2.4
120
GTM961200P12051-T3A*
51
2.35
119.85
GTM961200P12052-T3A*
52
2.3
119.6
GTM961200P12053-T3A*
53
2.26
119.78
GTM961200P12054-T3A*
54
2.22
119.88
GTM961200P10020-T3A*
20
6
120
GTM961200P120VV.V-T3A-XXXX
CUSTOM
CUSTOM
CUSTOM
* (Final ordering #/Part No to be defined)
Specifications
Input
Input Voltage:
Specified: 90-264VAC, Nameplate: 100-240VAC 100% rated load current for 90-264VAC 90% rated load current for 85-264VAC 100% rated load current for 90-370VDC
Input Frequency:
Specified: 47-63Hz, Nameplate: 50-60Hz
Power Factor:
0.90 min. @ 230VAC, 0.97 min. @ 115VAC
No Load Input Power:
< 150mW @ 115/230VAC
Inrush Current:
< 65A @ 115VAC, < 130A @ 230VAC (cold start)
Efficiency:
DoE Level VI and EU CoC Tier 2 compliant
Output
Turn-on Delay:
1 second typ.
Output Regulation:
± 4% (measured at output connector)
Line Regulation:
± 0.5% typ. (measured at output connector, full load)
Ripple:
1% or 200mV, whichever is greater (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW, at output connector)
Startup Overshoot:
5% max.
Transient Response:
5% max. deviation, 1ms typ. recovery time (with 25% to 50% load step)
Hold-up Time:
20ms min.
Power Indicator:
Green LED
Switching Frequency:
Variable, 25KHz - 125KHz, depending on load
Protections
Input Protection:
300VAC MOV transient suppressor, input line fusing
Over-Current Protection
Current limited to less than < 170W / Vout(nom.) and < 13A, auto-restart
Short-Circuit Protection:
Auto-restart
Over-Voltage Protection:
110 - 135% (60V max for models rated 45 - 54V), full load condition, latched off, cycle AC to reset
Over-Temperature Protection
Internal ambient NTC temperature sensor, latched off, cycle AC to reset
Complies with EU 2011/65/EU and China SJ/T 11363-2006
Note 1:
Regarding operating temperature a. See below de-rating table for output power capability at alternate temperature b. Extended low-end temperature range available as a custom option
Safety
Dielectric Withstand Voltage:
4000VAC or 5656VDC from input to output 3000VAC or 4242VDC from input to earth (Class I models only)
Touch Current:
Class I: 20µA max. Class II: 70µA max.
Earth Leakage Current
Class I: 300µA max. Class II: N/A
Means of Protection:
2 x MOPP
Output Isolation Options:
Class I: C6 or C14 inlet (output negative tied to Earth contact) Class II: C8 or C18 inlet Class II FE: C6 or C14 inlet (output isolated from Earth contact) Review isolation options by reading our white paper PSU Isolation and Identity.
Primary-to-Secondary Bridging Capacitors:
Two IEC60384-14 Class Y1 capacitors
Earth Continuity Test:
< 0.1Ω between earth pin (on AC inlet) and PCB termination point (Class I models only)
Compliant Standards:
See listing at end of this datasheet for specific standards
Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:
Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:
EN61000-3-2, Class A
Voltage Fluctuations/Flicker:
EN61000-3-3
Electrostatic Discharge (ESD) Immunity:
EN61000-4-2, 10KV contact discharge, 18KV air discharge
Radiated RF Immunity:
EN61000-4-3, 10V/m 80-1000MHz, 3V/m 1-2.7GHz, 80% 1KHz AM
EFT/Burst Immunity:
EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:
EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:
EN61000-4-6, 3VRMS, 80% 1KHz AM
Power Frequency Magnetic Field Immunity:
EN61000-4-8, 3A/m
Voltage Dip Immunity:
EN61000-4-11
Enclosure
Housing:
High impact plastic, 94V0 polycarbonate, non-vented
Size:
149.4 * 62.6 * 33.5mm, ± 1.0mm
AC Input Mechanical Options:
C6, C8, C14, or C18 IEC inlet
Markings:
Label and/or pad printed and/or laser-etched into case
Special Options
Reduced leakage current to meet IEC60601-1 Type-CF limits
Improved EN61000-4-5 line surge immunity (3KV differential, 5KV common-mode)
Reduced harmonic current levels for LED lighting (EN61000-3-2 Class C) for 85 - 120W load
Short term output surge capability for motors or other high peak current loads
Ruggedized construction with improved shock and vibration resistance
Compliance to IPC-600, IPC-610, IPC-6012 Class 2 or 3
Custom cordsets, special cordage and connector types
Custom housing markings
Custom housing/cordage colors
Tightened output voltage tolerance
Reduced output ripple level
Reduced output power rating
Low temperature option, down to -40°C
Epoxy potted version "P2" or "P3" suffix, with flying wires or custom cables
"Back EMF" upgrade, for motor loads and other high inductance applications with reverse energy flow requirements
Improved ingress protection (IP) rating
Output Connectors
Several output connector options are available with various output current ratings. GlobTek can supply 10A rated 2.1mm and 2.5mm style DC Power Jacks, to complement our 10A output rated 2.1mm and 2.5mm DC power plugs used on our output cordsets.
Below are standard cordsets which are "not included" (unless stated above); these can be purchased separately or package with the power supply. Contact your Sales Engineer if the style required is not shown below. Many more available in different lengths, colors or cable material.
Contact GlobTek for your specific requirements or custom solutions.
Approvals
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.
Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.
Logo
Description
Medical 4th Ed.
EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
Declaration ДС № ЕАЭС N RU Д-US.КА01.В.10453_19 Custom Union of Russia, Belarus and Kazakhstan http://www.globtek.com/redirect/?loc=gost-certificate-eac-declaration
Conforms to UL STD. 60950-1 Certified to CSA STD C22.2 NO.60950-1
Conforms to UL STD. 60950-1 Certified to CSA STD C22.2 NO.60950-1
Conforms to AAMI STD.ES60601-1,IEC 60601-1-11 Certified to CAN/CSA STD.C22.2 NO.60601-1
Conforms to AAMI STD.ES60601-1,IEC 60601-1-11 Certified to CAN/CSA STD.C22.2 NO.60601-1
RCM certificate GMA-514113-EA; Australia and New Zealand Regulatory Compliance, Mark (http://rcm.standards.org.au/rcmfaq/rcmfaq.htm)
CB for IEC61347-1:2015 + A1:2017 IEC61347-2-13:2014, A1:2016
Test standard: EN 55032: 2012+AC: 2013 EN 61000-3-2:2014 EN 61000-3-3:2013 EN 55024:2010 EN 60601-1-2:2015
Japan: Voluntary Control Council for Interference (VCCI)
Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (R2012) [AAMI ES60601-1:2005 +C1;A2] [CSA C22.2#60601-1:2014 Ed.3] [AAMI HA60601-1-11:2015 Ed.2]
No Logo Applicable
IEC 60335-1.2010+A1
CHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
No Logo Applicable
CB for IEC61347-1:2015 + A1:2017 IEC61347-2-13:2014, A1:2016
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
CCC Altitude up to 5000 m GB17625.1-2012, GB4943.1-2011, GB/T9254-2008
S-Mark Certificate EN 60601-1:2006 + A1:2013 + A12:2014 EN 60601-1-11:2015 (http://www.intertek.com/marks/s/)
RoHS
Specifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [UL 62368-1:2014 Ed.2] Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [CSA C22.2#62368-1:2014 Ed.2]
Indoor Use Only - Mark is on the label or Molded in the case
S-Mark Certificate EN 62368-1:2014 + A11 (http://www.intertek.com/marks/s/)
PRECAUCION: PARA USO EN EQUIPOS ELECTRONICOS SOLAMENTE
IRAM certificate DC-E-G71-005.1 to IEC 62368-1:2014
ETL (Brazil) to IEC 62368-1;2014 DOC cert
JAPAN TUV R-PSE, Cert. No. JD 50469658, to J62368-1(H30) , J55032(H29),J3000(H25)[DC30?60V]. Please reference the following website for guidelines on PSE regulations: https://www.globtek.com/r2/Szj4Vb
JAPAN TUV R-PSE, Cert. No. JD 50469658, to J62368-1(H30) , J55032(H29),J3000(H25)[15V or less]. Please reference the following website for guidelines on PSE regulations: https://www.globtek.com/r2/Szj4Vb
JAPAN TUV R-PSE, Cert. No. JD 50469658, to J62368-1(H30) , J55032(H29),J3000(H25)[DC15?30V]. Please reference the following website for guidelines on PSE regulations: https://www.globtek.com/r2/Szj4Vb