GTM96060-AWWVV-FW Open Frame/Internal Medical/ICT/ITE/Household use Power Supply

GTM96060-AWWVV-FW
GTM96060-AWWVV-FW, Medical/ICT/ITE/Household use Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, PANEL MOUNT IEC 320/C8, cable UL1007, 18AWG, Length  60±10 mm, White & Black, 
S&T 9±1 mm, Output Rating: 6 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals:GTM96060-AWWVV-FW, Medical/ICT/ITE/Household use Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, PANEL MOUNT IEC 320/C8, cable UL1007, 18AWG, Length 60±10 mm, White & Black, S&T 9±1 mm, Output Rating: 6 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals:
Information
Model NumberGTM96060-AWWVV-FW
Agency Documentshttp://www.globtek.info/certs/GTM96060/
CE EC-Declaration for MedicalContact GlobTek for more information
CE EC-Declaration for Industrial/Commercial/ITEContact GlobTek for more information
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for MedicalContact GlobTek for more information
UKCA-Declaration for Industrial/Commercial/ITEContact GlobTek for more information
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0O3a00000NLXAvEAP
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O3a00000NLXAvEAP
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryMedical/ICT/ITE/Household use Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)0.6 A
Max Output Power Rating (W)6
Output Voltage Range (V)5-48
EfficiencyN/A
Ingress ProtectionN/A
Size (mm)84.50 x 41.50 x 17.0
Enclosure




User-added image








 
3D File Downloadhttp://www.globtek.com/dl/GTM43007A-LOW&HIGH-VERSION.igs
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
GTM96060-A0606-1.0-FW*51.26
* (Final ordering #/Part No to be defined)
Specifications

Input

Input Voltage:Specified: 90-264VAC, Nameplate: 100-240VAC
     100% rated load current for 90-264VAC
       85% rated load current for 85-264VAC
     100% rated load current for 110-370VDC
Input Frequency:Specified: 47-63Hz, Nameplate: 50-60Hz
No Load Input Power:<75mW @ 230VAC
Inrush Current:<30A @ 230VAC (cold start @ 25°C)
Efficiency:DoE Energy Level VI and CoC Tier 2, for models 5V and above

Output

Turn-on Delay:1 second max. @ 100VAC and full load
Output Regulation± 5% (measured at the output connector)
Line Regulation:± 1%  (measured at the output connector @ full load)
Ripple:0.5% pk-pk or 50mV, whichever is greater (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW)
(Note: < 0.2% or 25mV option available upon request)
Transient Response:5% deviation max. (with 25% load step), recovery to within 1ms typ.
Startup Overshoot:<5% (full load)
Hold-up Time:8ms min. (100VAC and full load)
Power Indicator:Green LED
Switching Frequency:65KHz typ.

Protections

Input Protection:300Vac MOV transient suppressor, input line fusing
Over-Voltage Protection:110-130%, of nominal output voltage
Current Limiting:
(Standard CV Models):
Short circuit and overload protected, auto recovery
Output current limit: 120% to 250% of rated output current
(Upper end of current limit range is determined by normal output power rating, not derated output power level)
Current Limiting:
(Constant Current Models):
± 10% of specified nominal output current rating
Current regulated from Vout (nom.) to 0.25 * Vout (nom.)
Average output current below 0.25 * Vout will be reduced, and unit will hiccup cycle
Over-Temperature ProtectionOn-chip temperature sensor (integrated in primary-side switching controller), shut down at Tj = ~140°C, restart at Tj = ~70°C

Environmental

MTBF:1,500,000 hours @ 25°C ambient, full load (Telcordia SR-332, Issue 3)
Operating Temperature:-10°C to 50°C (@ 6W load)
See derating graph for output power capability outside of standard range.
Storage Temperature:-30°C to 80°C
Humidity:0% to 95% relative humidity, non-condensing
Cooling:Convection
RoHS:Compliant with latest regulations

Safety

Dielectric Withstand Voltage:4000VAC or 5656VDC from input to output
Touch Current:<60µA
Earth Leakage CurrentN/A, Class II input (No protective earth pin, therefore earth leakage is not measured)
Means of Protection:2 x MOPP
Primary-Secondary Bridging Capacitance:Two 470pF IEC60384-14 certified Y1 capacitors in series

EMC

Applicable Standards:Medical: EN60601-1-2, 4th edition
Emissions: EN55032, EN 55011, EN61000-6-3, EN61000-6-4, CISPR11, CISPR22
Immunity: EN55024, EN61000-6-1, EN61000-6-2
Conducted Emissions:Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:EN61000-3-2, Class A
Voltage Fluctuations/Flicker:EN61000-3-3
Electrostatic Discharge (ESD) Immunity:EN61000-4-2, 10KV contact discharge, 18KV air discharge
Radiated RF Immunity:EN61000-4-3, 10V/m 80-1000MHz, 3V/m 1-2.7GHz, 80% 1KHz AM
EFT/Burst Immunity:EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:EN61000-4-5, 1.5KV differential, 4KV common-mode
Conducted RF Immunity:EN61000-4-6, 3VRMS, 80% 1KHz AM
Power Frequency Magnetic Field Immunity:EN61000-4-8, 3A/m
Voltage Dip Immunity:EN61000-4-11

Enclosure

Housing:Open frame PCB
Size:84.5*41.5*17.0 +/-1.0mm
Markings:Label, pad printing, or laser engraving

Special Options

  1. Cost reduction, removal of LED
  2. Custom cordsets, various cordage types, and connector types
  3. Custom markings
  4. Reduced leakage current version, medical Type-CF leakage current compliance
  5. Tightened output voltage tolerance
  6. Reduced output ripple level
  7. Constant current battery charger characteristic
  8. Reduced output power rating
  9. High reliability PCB laminate with plated through-holes for IPC610 Class 2 Compliance
  10. Output voltage between 3.3V and 4.9V (without DoE EL6 and EU CoC Tier 2  compliance)
  11. Special housing colors and cordset colors
  12. Tamper proof, AC wall plug-in blade assembly. (Requires a special tool to remove the AC Wall Plug-in blade assembly.)
Derating CurveUser-added image
 
Qty (ea.)/Carton124
Individual Box/Gift Box Size (mm)95*65*35
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)442 x 330 x 275
Gross Weight (kg)15.2
Label Width (mm)25.0
Label Height (mm)17.0
Input Configuration
DescriptionPANEL MOUNT IEC 320/C8, cable UL1007, 18AWG, Length 60±10 mm, White & Black, S&T 9±1 mmUser-added image

Request Info